Temperature Calibration Equipment
01
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Multifunction Reference Thermometer Readout
Additel 286
Measure and calibrate SPRTs, RTDs, thermistors and thermocouples
1 PPM resistance ratio accuracy (channel 1)
8 1/2-digit DC multimeter
Measure up to 82 channels
Sample rates up to 10 channels per second
Bluetooth, WIFI, USB& Ethernet (RJ-45) capable
Auto temperature control of Additel and other manufacture's heat sources
Built-in automatic temperature control, data collection, and coefficient generation
Support for creating custom control of heat sources with RS-232
Auto - zero power feature (self heating compensation)
10.1 in. touch screen display
Supports fully automated temperature calibrations with data collection and report
generation (no software required)
OVERVIEW
The Additel 286 Multifunction Reference Thermometer Readout is an industry first!
We have combined the capabilities of a high-end reference thermometer with a highly
capable data acquisition system and 8.5 digit multimeter. The ADT286 is capable
of scanning and recording up to 82 channels at 10 channels per second. Users can
easily congure the ADT286 to perform eld calibrations and uniformity studies as well
as use the unit in the lab as a precision thermometer and 8.5 digit multimeter. Get
more for less with this newest game changer from Additel!
02
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
ADT 286 Multifunction Reference Thermometer Readout Scanner Modules
If you're in need of a precision reference thermometer for your laboratory, then look no further than the Additel 286.
The base unit comes with two precision readout channels that can be used to measure your SPRT. Need to calibrate
RTDs, PRTs, thermistors or thermocouples? Add a scanner module and you now have the ability to measure 10
RTDs, PRTs, or thermistors and 20 thermocouples. Expand up to 82 channels with our unique easy to use scanner
modules. Each 20 channel module is outfitted with our proprietary universal terminals with an industries best cold
junction capability second to none. Utilize the module docked atop the ADT286, or connect remotely with cables
to suit nearly any unique setup/configuration. Additel also has a process module specifically designed to measure
process instrumentation like transmitters and switches. This scanner will also supply loop power for the transmitters.
Designed to make your job easier, the ADT286 has a large sensor library supporting 15 TC types, both standard
and special limits, 18 different thermal resistors, CVD, ITS-90, and a large variety of standard curves for RTDs and
thermistors. The ADT286 is loaded with special applications such as probe calibration, SPRT calibration, chamber
mapping and more. And we continue to add applications on a regular basis!
Available in 25 & 100 Ohm values, users can enjoy improved resistance ratio performance by easily plugging one of our
reference resisters into channel 2 on the new ADT286. Perfect for calibrating your SPRT’s and high end PRT’s.
ADT286-RS Resistance Standards
Automatic Temperature Control and Probe Calibration
The Additel 286 Multifunction Reference Thermometer Readout has preinstalled drivers to control Additel and other
manufacturer’s heat sources. Simply connect to one or more heat sources via a communication cable, Ethernet or
wireless and now it will automatically control to the set point and desired stability. If your heat source is not on the list,
you can easily add the driver yourself so you can run automated calibrations with any heat source.
Now combine the heat source control feature with our probe calibration application and you have a very powerful
automatic calibration solution. The probe calibration app allows you to automatically setup and run calibration routines
with multiple set points and multiple heat sources, collect data, and develop calibration coefcients — all with one device
and without the need of software! Simply place a batch of sensors of any mix and type into your heat source, connect
it to the ADT286, run the probe calibration app and come back to a completed test. All that is left to do is generate
and export all the calibration data. There’s no need to work with complicated software for communication, set up or
coefcient generation. There is no reason to have a calibration technician manually monitor the process and record the
data. This Multifunction Reference Thermometer Readout will do all the work for you.
Temperature Calibration Equipment
03
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
FEATURES
Specification Display Application
Multi-Channel
Smart Connection
DAQ Mode
Temperature Mapping Mode
Probe Calibration
Environmental Temperature
Field Test
04
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
APPLICATIONS
Specification Display Application
SPRT Mode
Switch Testing
Thermostatic Source Test
DMM
Heat Source Control
Snapshot
Temperature Calibration Equipment
05
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
SPECIFICATIONS
Specification Description
Voltage
100V Setting 90V to 110V
120V Setting 108V to 132V
220V Setting 198V to 242V
240V Setting 216V to 264V
Frequency 47Hz to 440Hz.Automatically sensed at power-on
Power Consumption 40VA peak (30Watt average)
Temperature
Operating : 0°C to 50°C
Full accuracy :18°C to 28°C
Storage : -20°C to 70°C
Warm-up 90 mins for full uncertainty specications
Relative Humidity
(non-condensing)
Operating : 0°C to 28°C < 90%
28°C to 40°C < 75%
40°C to 50°C < 50%
Storage : -20°C to 70°C < 95%
Altitude
Operating : 2000 m
Storage : 12000 m
Vibration and Shock Complies with MIL-28800F Class 3
Input Protection 50V all functions, ranges and terminals
Communication
USB-A , USB-B , RJ45 , WiFi , Bluetooth
Memory 10G - All data stored with time stamps
Localization English , Chinese
Display 10.1 in (256 mm) TFT color display
Size (H x W x D) 9.8 in (250 mm) x 16.5 in (420 mm) x 7.9 in (200 mm)
Weight 18.5 lb ( 8.39 kg)
Other Conformities CE
General Specifications
Measurement Specifications
Specication Conditions: 90 mins Warm-Up Time / Environment Temperature (18 - 28) °C.
The following specications apply for front panel, after at least 90 minutes warm-up.
24-hour specications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326.
Resistance Ratio Accuracy (Rx/Rs) using External Rs
Range Reference Resistance Ratio (Rx/Rs)
1 Year (23
± 5) °°C
ppm of Reading
100 25 Ω
2.00-4.00 1.5
1.10-2.00 0.85
0.90-1.10 0.6
0.50-0.90 1.5
0.25-0.50 2.5
400 100 Ω
2.00-4.00 2
1.10-2.00 0.81
0.90-1.10 0.26
0.50-0.90 0.95
0.25-0.50 1.2
06
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Measurement
Range
Scanning Speed Resolution
24 Hour
(23
±1) °°C
90 Days
(23
± 5) °C
1 year
(23
± 5) °°C
Excitation
Currert
Temperature
Coefficient
( 0~100 )
Slow Speed 0.01 mΩ 3 ppm or 0.2 mΩ 12 ppm or 0.35 mΩ 15 ppm or 0.35 mΩ
±1 mA/±12 V 3 ppm + 0.01 mΩMedium Speed 0.01 mΩ 3 ppm or 0.55 mΩ 12 ppm or 0.7 mΩ 15 ppm or 0.7 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.7 mΩ 12.6 ppm or 1.85 mΩ 15.6 ppm or 1.85 mΩ
( 0~400 )
Slow Speed 0.01 mΩ 3 ppm or 0.3 mΩ 12 ppm or 0.4 mΩ 15 ppm or 0.4 mΩ
±1 mA/±12 V 3 ppm + 0.02 mΩMedium Speed 0.01 mΩ 3 ppm or 0.7 mΩ 12 ppm or 0.8 mΩ 15 ppm or 0.8 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.9 mΩ 12.6 ppm or 2 mΩ 15.6 ppm or 2 mΩ
( 0~4000 )
Slow Speed 0.1 mΩ 3 ppm or 4 mΩ 12 ppm or 5 mΩ 15 ppm or 5 mΩ
±0.1 mA/±12 V 3 ppm + 0.2 mΩ
Medium Speed 0.1 mΩ 3 ppm or 8 mΩ 12 ppm or 9 mΩ 15 ppm or 9 mΩ
Fast Speed 1 mΩ 3.6 ppm or 20 mΩ 12.6 ppm or 21 mΩ 15.6 ppm or 21 mΩ
Resistance Accuracy using Internal Rs
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefcient index: exceeds (18-28) °C range, increase per degree (ppm reading +xxmΩ).
[3] Specications are for 4-wire function. For 3-wire, add 0.005 Ω for internal resistance mismatch. For 2-wire, add 0.005 Ω for internal resistance
[4] Automatic current reversal.
Scanning Speed Temperature
24 Hour /
°C
(23
±1) °°C
90 Days /
°C
(23
± 5) °°C
1 year /
°C
(23
± 5) °°C
Temperature Coefficient
°C/°C
Slow Speed
-200 °C 0.0005 0.0008 0.0008 0.0002
0 °C 0.0008 0.0031 0.0038 0.0008
300 °C 0.0018 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.0030
Medium Speed
-200 °C 0.0013 0.0016 0.0016 0.0002
0 °C 0.0014 0.0031 0.0038 0.0008
300 °C 0.0020 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.0030
Fast Speed
-200 °C 0.0039 0.0043 0.0043 0.0006
0 °C 0.0044 0.0047 0.0047 0.0013
300 °C 0.0053 0.0093 0.0093 0.0024
600 °C 0.0059 0.0152 0.0152 0.0036
PRT Measurement Accuracy using Internal Rs
[1] The indicator is based on the electrical accuracy of the 4-wire PT100 PRT and does not include the accuracy of the PRT itself.
[2] Temperature maximum Resolution is 0.0001 °C.
SPRT/PRT Measurement Accuracy using External Rs
SPRT/PRT Type
External Reference
Resistance
Temperature (
°C) °
Resistance Ratio
(Rx/Rs)
1 Year(23
± 5) °°C
ppm of reading
Equivalent to
Temperature (mK)
PT25 25 Ω
-189.3442 0.22 2.5 0.13
-38.8344 0.84 1.5 0.32
0.01 1 0.6 0.15
231.928 1.89 0.85 0.44
419.527 2.57 1.5 1.11
660.323 3.37 1.5 1.58
PT100 100 Ω
-189.3442 0.22 1.2 0.07
-38.8344 0.84 0.95 0.20
0.01 1 0.26 0.07
231.928 1.89 0.81 0.42
419.527 2.57 2 1.47
660.323 3.37 2 2.11
[1] The PT25 indicator is based on a nominal resistance of 25 Ω for Rx.
[2] The PT100 indicator is based on a nominal resistance of 100 Ω for Rx.
[3] The uncertainty of external Rs is not included. The user may choose the ADT280-RS-25/100 standard resistor as external Rs, which has an accuracy
of 5 ppm at (23±2) °C.
Temperature Calibration Equipment
07
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Thermocouple Cold Junction Accuracy
CJC Accuracy
±0.1
°C , 1 year, 23 °C
± 5
°C
Environmental Coefcient Beyond (18 ~ 28) °C, add 0.02 °C / °C
Other Each signal scanner has 10 cold Junction temperature sensors
Thermocouple Temperature Accuracy
Type Temperature
24 Hour/
°°C
(23
±1) °°C
90 days/
°°C
(23
± 5) °°C
1 year /
°°C
(23
±5) °°C
Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed
E
-200 0.089 0.038 0.022 0.099 0.047 0.031 0.100 0.049 0.033
-100 0.049 0.021 0.012 0.054 0.026 0.017 0.055 0.026 0.017
-40 0.041 0.017 0.009 0.045 0.021 0.013 0.045 0.021 0.014
0 0.038 0.015 0.009 0.041 0.019 0.012 0.041 0.019 0.012
155 0.031 0.013 0.008 0.035 0.017 0.011 0.036 0.017 0.012
350 0.029 0.013 0.008 0.033 0.017 0.012 0.035 0.018 0.013
660 0.031 0.014 0.009 0.036 0.020 0.015 0.039 0.022 0.017
1000 0.034 0.017 0.012 0.042 0.025 0.019 0.046 0.029 0.024
J
-200 0.102 0.043 0.025 0.113 0.054 0.036 0.115 0.055 0.037
-100 0.054 0.022 0.013 0.060 0.028 0.018 0.060 0.028 0.019
-40 0.047 0.019 0.011 0.051 0.024 0.015 0.051 0.024 0.015
0 0.044 0.018 0.010 0.048 0.022 0.014 0.048 0.022 0.014
155 0.041 0.017 0.010 0.045 0.021 0.014 0.046 0.022 0.015
350 0.042 0.018 0.011 0.047 0.023 0.016 0.048 0.025 0.018
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.020
1200 0.044 0.022 0.015 0.054 0.031 0.024 0.059 0.036 0.029
K
-200 0.146 0.061 0.035 0.161 0.076 0.050 0.163 0.077 0.051
-100 0.073 0.030 0.017 0.080 0.037 0.024 0.080 0.038 0.025
-40 0.060 0.025 0.014 0.066 0.030 0.020 0.066 0.031 0.020
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.056 0.023 0.013 0.061 0.029 0.019 0.062 0.030 0.020
350 0.054 0.023 0.014 0.061 0.030 0.020 0.062 0.031 0.021
660 0.055 0.025 0.015 0.063 0.033 0.023 0.066 0.035 0.026
1372 0.073 0.035 0.023 0.087 0.049 0.037 0.093 0.055 0.043
Thermocouple Voltage Accuracy
Test Range Scanning Speed Resolution
24 hours
(23
±1) °°C
90 days
(23
±5) °°C
1 year
(23
±5) °°C
Input Resistance
Temperature
Coefficient
(-100-100) mV
Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm
10 MΩ or >10 GΩ 1 ppm + 0.1 µV
Medium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature coefcient index: Exceed the range of (18-28) °C, increase (ppm of reading + xxμV)/°°C.
08
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Type Temperature
24 Hour/
°°C
(23
±1) °°C
90 days/
°°C
(23
± 5) °°C
1 year /
°°C
(23
±5) °°C
Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed
T
-200 0.142 0.059 0.034 0.156 0.073 0.048 0.157 0.075 0.049
-100 0.078 0.032 0.018 0.086 0.040 0.026 0.086 0.040 0.026
-40 0.063 0.026 0.015 0.069 0.032 0.020 0.069 0.032 0.021
0 0.057 0.023 0.013 0.062 0.028 0.018 0.062 0.028 0.018
155 0.044 0.019 0.011 0.049 0.023 0.015 0.049 0.024 0.016
350 0.038 0.016 0.010 0.043 0.021 0.015 0.044 0.022 0.016
400 0.037 0.016 0.010 0.042 0.021 0.015 0.044 0.023 0.016
R
-40 0.543 0.222 0.124 0.593 0.272 0.173 0.593 0.272 0.173
0 0.416 0.170 0.095 0.454 0.208 0.132 0.454 0.208 0.132
155 0.266 0.109 0.061 0.290 0.134 0.086 0.291 0.134 0.086
350 0.220 0.091 0.051 0.241 0.112 0.072 0.242 0.113 0.073
660 0.192 0.080 0.046 0.212 0.100 0.066 0.214 0.102 0.068
1768 0.188 0.082 0.049 0.213 0.107 0.074 0.219 0.114 0.081
S
-40 0.515 0.211 0.117 0.562 0.258 0.164 0.562 0.258 0.164
0 0.407 0.167 0.093 0.444 0.204 0.130 0.444 0.204 0.130
155 0.275 0.113 0.063 0.300 0.138 0.089 0.301 0.139 0.089
350 0.236 0.098 0.055 0.259 0.120 0.078 0.260 0.122 0.079
660 0.214 0.089 0.051 0.236 0.111 0.073 0.239 0.114 0.075
1768 0.222 0.096 0.057 0.250 0.124 0.086 0.257 0.132 0.093
B
250 0.872 0.357 0.199 0.952 0.437 0.278 0.952 0.437 0.279
350 0.619 0.254 0.141 0.676 0.311 0.198 0.676 0.311 0.199
660 0.342 0.141 0.079 0.374 0.173 0.111 0.375 0.175 0.113
1820 0.199 0.085 0.050 0.222 0.108 0.073 0.227 0.113 0.078
N
-200 0.224 0.093 0.052 0.246 0.115 0.075 0.247 0.116 0.076
-100 0.106 0.044 0.024 0.116 0.054 0.035 0.116 0.054 0.035
-40 0.089 0.036 0.020 0.097 0.045 0.029 0.097 0.045 0.029
0 0.084 0.035 0.019 0.092 0.042 0.027 0.092 0.042 0.027
155 0.070 0.029 0.017 0.077 0.036 0.024 0.078 0.037 0.024
350 0.062 0.026 0.015 0.069 0.033 0.022 0.070 0.035 0.024
660 0.059 0.026 0.016 0.067 0.034 0.024 0.069 0.036 0.026
800 0.060 0.027 0.016 0.068 0.035 0.025 0.071 0.038 0.028
1000 0.062 0.028 0.018 0.072 0.038 0.028 0.075 0.042 0.031
1200 0.065 0.030 0.019 0.076 0.041 0.031 0.081 0.046 0.035
1300 0.068 0.032 0.020 0.080 0.044 0.033 0.085 0.049 0.038
L
-200 0.069 0.029 0.017 0.076 0.036 0.024 0.077 0.037 0.025
-100 0.053 0.022 0.013 0.059 0.028 0.018 0.059 0.028 0.018
-40 0.045 0.019 0.010 0.049 0.023 0.015 0.050 0.023 0.015
0 0.043 0.018 0.010 0.047 0.021 0.014 0.047 0.021 0.014
155 0.040 0.017 0.010 0.044 0.021 0.014 0.045 0.022 0.015
350 0.041 0.018 0.011 0.046 0.023 0.016 0.047 0.024 0.017
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.020
900 0.035 0.017 0.011 0.042 0.023 0.017 0.045 0.026 0.021
U
-80 0.072 0.030 0.017 0.079 0.037 0.024 0.079 0.037 0.024
-40 0.062 0.026 0.014 0.068 0.031 0.020 0.068 0.032 0.020
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.045 0.019 0.011 0.049 0.023 0.015 0.050 0.024 0.016
350 0.037 0.016 0.010 0.042 0.021 0.014 0.043 0.022 0.016
600 0.034 0.015 0.010 0.039 0.021 0.015 0.041 0.023 0.017
[1] The index is based on the accuracy of the thermocouple electrical measurement of temperature scanner module, does not include the accuracy of the
thermocouple itself and the xed cold junction compensation at 0 °C.
[2] The highest temperature resolution is 0.0001 °C.
Thermocouple Temperature Accuracy
Temperature Calibration Equipment
09
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
DC Voltage Accuracy
Test Range Scanning Speed Resolution
24 hours
(23
±1) °°C
90 days
(23
±5) °°C
1 year
(23
±5) °°C
Input Resistance
Temperature
Coefficient
(-100-100) mV
Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm
>10 GΩ or 10 MΩ 1 ppm + 0.1 µVMedium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
(-1-1) V
Slow Speed 0.1 µV 2 ppm + 0.3 ppm 8 ppm + 0.6 ppm 14 ppm + 0.6 ppm
>10 GΩ or 10 MΩ 1 ppm + 0.2 µVMedium Speed 0.1 µV 2 ppm + 1.3 ppm 8 ppm + 1.6 ppm 14 ppm + 1.6 ppm
Fast Speed 1 µV 2.6 ppm + 3.3 ppm 8.6 ppm + 3.6 ppm 14.6 ppm + 3.6 ppm
(-10-10) V
Slow Speed 1 µV 2 ppm + 0.05 ppm 8 ppm + 0.08 ppm 14 ppm + 0.08 ppm
>10 GΩ or 10 MΩ 1 ppm + 0.3 µVMedium Speed 1 µV 2 ppm + 0.35 ppm 8 ppm + 0.38 ppm 14 ppm + 0.38 ppm
Fast Speed 10 µV 2.6 ppm + 1.05 ppm 8.6 ppm + 1.08 ppm 14.6 ppm + 1.08 ppm
(-50-50) V
Slow Speed 10 µV 8 ppm + 1 ppm 32 ppm + 1 ppm 38 ppm + 1 ppm
10 MΩ 5 ppm + 5 µVMedium Speed 10 µV 8 ppm + 2 ppm 32 ppm + 2 ppm 38 ppm + 2 ppm
Fast Speed 100 µV 8.6 ppm + 7 ppm 32.6 ppm + 7 ppm 38.6 ppm + 7 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index: Exceed the range of (18-28) °C, increase (ppm reading + xxμV)/°°C.
[3] Any range, the maximum input voltage is 50 V.
Measurement
Range
Scanning Speed Resolution
24 Hour
(23
±1) °°C
90 Days
(23
± 5) °C
1 year
(23
± 5) °°C
Excitation
Currert
Temperature
Coefficient
( 0~12 ) k
Slow Speed 1 mΩ 10 ppm or 60 mΩ 30 ppm or 80 mΩ 40 ppm or 80 mΩ
10 µA 5 ppm + 10 mΩMedium Speed 1 mΩ 10 ppm or 110 mΩ 30 ppm or 130 mΩ 40 ppm or 130 mΩ
Fast Speed 10 mΩ 10 ppm or 210 mΩ 30 ppm or 230 mΩ 40 ppm or 230 mΩ
( 10~120 ) k
Slow Speed 10 mΩ 10 ppm 30 ppm 40 ppm
10 µA 5 ppm + 20 mΩMedium Speed 10 mΩ 10 ppm + 80 mΩ 30 ppm + 80 mΩ 40 ppm + 80 mΩ
Fast Speed 100 mΩ 10.6 ppm + 200 mΩ 30.6 ppm + 200 mΩ 40.6 ppm + 200 mΩ
( 100~1000 ) k
Slow Speed 0.1 Ω 50 ppm 80 ppm 100 ppm
1 µA 5 ppm + 1 ΩMedium Speed 0.1 Ω 50 ppm + 1 Ω 80 ppm + 1 Ω 100 ppm + 1 Ω
Fast Speed 1 Ω 51 ppm + 2 Ω 81 ppm + 2 Ω 101 ppm + 2 Ω
Thermistor Accuracy
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefcient index: exceeds (18-28) °C range, increase (ppm reading +xxmΩ) / °C.
[3] Specications are for 4-wire function.
Type Scanning Speed Temperature
24 Hour /
°C
(23
±1) °°C
90 Days /
°C
(23
± 5) °°C
1 year /
°C
(23
± 5) °°C
10 k
Slow Speed
-40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0004 0.0008 0.0011
100 °C 0.0030 0.0039 0.0039
150 °C 0.0130 0.0174 0.0174
Medium Speed
-40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0008 0.0010 0.0011
100 °C 0.0054 0.0064 0.0064
150 °C 0.0239 0.0282 0.0282
Fast Speed
-40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0016 0.0016 0.0016
100 °C 0.0104 0.0104 0.0104
150 °C 0.0456 0.0456 0.0456
Thermistor Temperature Accuracy
[1] The indicator is based on the electrical accuracy of the 4-wire thermistor and does not include the accuracy of the thermistor itself.
[2] Temperature maximum Resolution is 0.0001 °C.
10
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
DC Current Accuracy
Test Range Scanning Speed Resolution
24 hours
(23
±1) °°C
90 days
(23
±5) °°C
1 year
(23
±5) °°C
Burden
Voltage
Temperature Coefficient
(-100-100) µA
Slow Speed 0.01 nA 15 ppm + 3 ppm 50 ppm + 6 ppm 60 ppm + 6 ppm
<1 mV 8 ppm + 0.1 nAMedium Speed 0.01 nA 15 ppm + 7 ppm 50 ppm + 10 ppm 60 ppm + 10 ppm
Fast Speed 0.1 nA 15 ppm + 23 ppm 50 ppm +26 ppm 60 ppm + 26 ppm
(-1-1) mA
Slow Speed 0.1 nA 15 ppm + 0.6 ppm 50 ppm +1 ppm 60 ppm + 1 ppm
<1 mV 8 ppm + 0.5 nA
Medium Speed 0.1 nA 15 ppm + 1.6 ppm 50 ppm + 2 ppm 60 ppm + 2 ppm
Fast Speed 1 nA 15.6 ppm + 3.6ppm 50.6 ppm + 4 ppm 60.6 ppm + 4 ppm
(-10-10) mA
Slow Speed 1 nA 30 ppm + 3 ppm 75 ppm + 6 ppm 80 ppm + 6 ppm
<1 mV 8 ppm + 10 nA Medium Speed 1 nA 30 ppm + 7 ppm 75 ppm + 10 ppm 80 ppm + 10 ppm
Fast Speed 10 nA 30 ppm + 23 ppm 75 ppm + 26 ppm 80 ppm + 26 ppm
(-100-100) mA
Slow Speed 10 nA 40 ppm + 0.6 ppm 75 ppm + 1 ppm 80 ppm + 1 ppm
<1 mV 8 ppm + 50 nA Medium Speed 10 nA 40 ppm + 1.6 ppm 75 ppm + 2 ppm 80 ppm + 2 ppm
Fast Speed 100 nA 40.6 ppm + 3.6 ppm 75.6 ppm + 4 ppm 80.6 ppm + 4 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xxnA)/°°C.
[3] Input Protection 0.3A/600V Resettable PTC.
DC Resisstance Accuracy
Test Range Scanning Speed Resolution
24 hours
(23
±1) °°C
90 days
(23
±5) °°C
1 year
(23
±5) °°C
Excitation
Current
Temperature Coefficient
(0-100)
Slow Speed 0.01 mΩ 3 ppm + 1 ppm 13 ppm + 1.5 ppm 16 ppm + 1.5 ppm
1 mA 3 ppm + 0.01 mΩ Medium Speed 0.01 mΩ 3 ppm + 5 ppm 13 ppm + 5.5 ppm 16 ppm + 5.5 ppm
Fast Speed 0.1 mΩ 3 ppm + 21 ppm 13 ppm + 21.5 ppm 16 ppm + 21.5 ppm
(0-1) k
Slow Speed 0.1 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm
1 mA 3 ppm + 0.02 mΩMedium Speed 0.1 mΩ 3 ppm + 1.2 ppm 12 ppm + 1.3 ppm 15 ppm + 1.3 ppm
Fast Speed 1 mΩ 3.6 ppm + 3.2 ppm 12.6 ppm + 3.3 ppm 15.6 ppm + 3.3 ppm
(0-10) k
Slow Speed 1 mΩ 3 ppm + 0.3 ppm 12 ppm + 0.4 ppm 15 ppm + 0.4 ppm
0.1 mA 3 ppm + 0.2 mΩMedium Speed 1 mΩ 3 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 15 ppm + 1.3 ppm
Fast Speed 10 mΩ 3.6 ppm + 3.3 ppm 12.6 ppm + 3.4 ppm 15.6 ppm + 3.4 ppm
(0-100) k
Slow Speed 10 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm
0.1 mA 3 ppm + 20 mΩMedium Speed 10 mΩ 3 ppm + 0.5 ppm 12 ppm + 0.6 ppm 15 ppm + 0.6 ppm
Fast Speed 100 mΩ 3.6 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 30.6 ppm + 1.3 ppm
(0-1) M
Slow Speed 0.1 Ω 10 ppm + 0.6 ppm 30 ppm + 1 ppm 40 ppm + 1 ppm
10 µA 5 ppm + 0.2 ΩMedium Speed 0.1 Ω 10 ppm + 1.2 ppm 30 ppm + 0.6 ppm 40 ppm + 0.6 ppm
Fast Speed 1 Ω 10 ppm + 2.6 ppm 30 ppm + 3 ppm 40 ppm + 3 ppm
(0-10) M
Slow Speed 1 Ω 50 ppm + 0.4 ppm 80 ppm + 1 ppm 100 ppm + 1 ppm
1 µA 10 ppm + 1 Ω Medium Speed 1 Ω 50 ppm + 1.4 ppm 80 ppm + 2 ppm 100 ppm + 2 ppm
Fast Speed 10 Ω 50 ppm + 4.4 ppm 80 ppm + 5 ppm 100 ppm + 5 ppm
(0-100) M
Slow Speed 10 Ω 150 ppm + 1 ppm 400 ppm + 4 ppm 500 ppm + 4 ppm
0.1 µA 50 ppm + 50 ΩMedium Speed 10 Ω 150 ppm + 6 ppm 400 ppm + 9 ppm 500 ppm + 9 ppm
Fast Speed 100 Ω 150 ppm + 11 ppm 400 ppm + 14 ppm 500 ppm + 14 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xx Ω)/°°C.
[3] The above is a 4-wire measurement index.
[4] When the range is less than or equal to 10 kΩ, the default is automatic current reversal.
[5] Max Lead Resistance(4-wire ohms): 10 Ω per lead for 100 Ω & 1 kΩ ranges; 100 Ω per lead for 10 kΩ &100 kΩ ranges; 1 kΩ per lead on all other ranges.
Temperature Calibration Equipment
11
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Accessories (Included)
Standard Accessories Quantity Picture
Shorting Block
1210103531
1 pc.
USB Cable (UK-415)
1210200243
1 pc.
Test leads
4 sets (8 pcs)
9026
2-wire test lead
(Only w/ ADT286-TS-PKG & ADT286-
PS-PKG)
20 pcs
Fuse (0213.315MXP)
for 110V units
2 pcs
Fuse (0218.160MXP)
(for 220V units)
2 pcs
ISO 17025 Accredited calibration 1 pc.
CD Manual 1 pc.
Ordering Information
Accessories
Model Number
Model Description Picture
ADT286-110V
ADT286-220V
Multifunction Reference
Thermometer Readout base
unit only
ADT286-TS-PKG-
110V
ADT286-TS-PKG-
220V
Multifunction Reference
Thermometer Readout base
unit with (1) Temperature
Scanner Module (9051 cable
not included)
ADT286-PS-PKG-
110V
ADT286-PS-PKG-
220V
Multifunction Reference
Thermometer Readout base
unit with (1) Process Scanner
Module (9051 cable not
included)
Optional Accessories
Model
Optional
Accessories
Picture
9026 2-wire test lead (20-Pack)
9051-10 Dsub Comm Cable=10 ft
9051-33 Dsub Comm Cable=33 ft
ADT286-DOCK
Remote Module Docking
Station w/AC Adapter
ADT286-TS
ADT286 Temperature
Scanner Module
ADT286-PS
ADT286 Process Scanner
Module
ADT280-RS-25
25 Ω Standard Reference
Resistor
ADT280-RS-100
100 Ω Standard Reference
Resistor
9916-286
Carrying Case for
ADT286,(2) scanner
modules and
reference probe w/
wheels
1pc
Secondary Standard PRT Ordering Information
AM1760
SP
12
Secondary PRT Model:
AM1760
AM1762
PRT Exterior:
12 - 12 inch straight
20 - 20 inch straight
AM17XX-X-SP
Resistance Standards
Specification ADT280-RS-25 ADT280-RS-100
Nominal
Resistance
25 Ω 100 Ω
Stability 5 ppm/year 5 ppm/year
Operating
Temperature
23 °C±2 °C 23 °C±2 °C
Temperature
Coefcient
0.5 ppm/°C 0.5 ppm/°C
Size
57 mm x 57 mm x 45 mm 57 mm x 57 mm x 45 mm
Weight 0.35 lb (160 g) 0.35 lb (160 g)
Excitation Current 1 mA 1 mA
ADT280-RS
25
Norminal Resistance:
25 - 25
100 - 100
ADT280-RS-X
Standard Reference Resistor Ordering Information
12
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20240321
Specification AM1760 Series AM1762 Series
Temperature Range -200 °C to 670 °C -200 °C to 670 °C
Resistance at 0
Nominal 100 Ω Nominal 25 Ω
Temperature Coefficient 0.003925 Ω / Ω / °C
Accuracy
±
0.007 °C at -196 °C
±
0.006 °C at 0.01 °C
±
0.015 °C at 420 °C
±
0.025 °C at 660 °C
±
0.007 °C at -196 °C
±
0.006 °C at 0.01 °C
±
0.015 °C at 420 °C
±
0.025 °C at 660 °C
Drift
±
0.004 °C at TPW after 100 hours at 661 °C
Short Term Stability
±
0.002 °C
Thermal Shock
±
0.002 °C after 10 times thermal cycles from minimum to maximum temperatures
Hysteresis N/A
Self-heating 0.0015 °C at 1 mA current
Response Time 9 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current 0.5 mA or 1 mA
Sensor Length 42 mm
Sensor Location 5 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconel
tm
Dimension
AM1760-12-SP
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1760-20-SP
0.25 in dia X 20 in (6.35 mm X 500 mm)
AM1762-12-SP
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1762-20-SP
0.25 in dia X 20 in (6.35 mm X 500 mm)
External Leads Teon
tm
– insulated copper wire, 4 leads, 2.5 meters
Handle Dimension 15 mm (OD) x 65 mm (L)
Handle Temperature Range
[1]
-50 °C to 160 °C -50 °C to 180° C
Calibration NIST traceable calibration w/ data included
[1] Handle temperatures outside the usable will cause damage to the probe.
* PRT Information from www.accumac.com.
Secondary Standard PRT Information