Pressure Calibration Equipment
Visit our website at www.additel.com or call 714-998-6899
02
Model ADT680 ADT680W
Wireless Communication
(ADT680W only)
N/A
Wireless Frequency: 2.4 G ISM Bands, 20 meter range
Number of wireless Channels : Chanel 1-15
Software: Wireless network demo software included read up
to 20 gauges.
Data Logging
(ADT680W only)
N/A
Storage Capacity: 140,000 readings (time, pressure, and
temperature)
Storage Interval: Adjustable from 1-9999 Sec
Single-button-press data logging enabled
Key Lockout: When the gauge is in auto-storing mode, the
front panel buttons will be automatically locked.
Filtering
Averaging (3 to 10 samples) or low-pass rst-order lter.
Max/Min data capture
Saves Max and Min data during pressure measurement.
Pressure units
Pa, kPa, MPa, bar, mba, psi, kgf/cm
2
, mmH
2
O, mmHg, inH
2
O, inHg, ozf/in, %, °C, °F
Engineering units: inH
2
O(20°C), inH
2
O(60°F), mmH
2
O(20°C), mmH
2
O(15°C), ftH
2
O(60°F), or ftH
2
O(4°C)
Display
LCD Specication: FSTN-LCD,Visual scope 36x61 mm
Full 5 digits,15.2 mm High
7 segment analog bar graph scaled from 0-100% of FS
Backlight: White
Backlight Duration: Not auto off,15,30,45,60 seconds optional
Auto off
Disabled,15, 30, 45, 60, 90, or 120 Minutes
Compliance
Certicates: CE R&TTE, FCC ID, IC ID
Protection Level: IP67
Vibration: 5 g (20-2000Hz)
Shock resistance: 100 g/11 ms
Pressure Port
≤15,000 PSI: 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP male, M20×1.5 male
>15,000 PSI: 1/4HP female or 1/4HP male
*1/4HP female: Autoclave F-250-C, 9/16" - 18 UNF-2B
*1/4HP male: Autoclave M-250-C, 9/16" - 18 UNF-2A
Other connections available per request.
Overpressure Alarm
Display will ash over 120%FS
Battery voltage Indicator
Displays the battery life remaining. When the battery voltage is too low, the gauge will power-off automatically.
Battery Life
1500 hours (10 readings/sec), 3000 hours (3 readings/sec), 6000 hours (1 readings/sec), 12000 hours (1 readings/15 s)
(100 hours when wireless communication is on)
Overpressure Record
Gauge will record max pressure data when the gauge is over pressured by 120% of FS.
Leakage test
In leak test mode, the gauge will record beginning pressure, ending pressure, and show the difference ΔP.
Factory Reset
Resets all settings back to factory default, except the calibration parameters.
Warranty
1 year
SPECIFICATIONS
Compound Pressure
P/N
Pressure Range
Accuracy(FS%)
[2]
Media
[3]
Burst Pressure
(psi)
[1]
(bar)
CP15 ±15 ±1 0.05(0.1, 0.25) G 3×
CP30 -15 to 30 -1 to 2 0.05(0.1, 0.25) G 3×
[1] Sealed gauge pressure for above 1,000 psi.
[2] FS specication applies to the span of the range.
[3] G=Gas
PRESSURE RANGE